Calibration vs. Adjustment

Calibration vs. Adjustment 

Q-Lab strongly recommends regular calibration of irradiance sensors, black panel and chamber air temperature sensors, and relative humidity probes on our testers. This ensures the most consistent and accurate tester operation. Often, however, the term calibration is used when the proper term should be adjustment, or a combination of the two terms. The difference can be important for auditing and quality system purposes.

A calibration procedure is one that involves a comparison to a known measurement standard. Often that standard is traceable back to a national metrological institute, for example NIST in the United States. An adjustment, on the other hand, is a procedure wherein a change is made to a measuring system so that it outputs proper values of the quantity to be measured. Although these procedures are often performed together, one does not require the other. A simple example to understand this framework is setting your wristwatch.

1) You can compare the time on the watch to the official United States time, at time.gov. If the watch reads 10:55 and time.gov says 11:00, then your watch has been compared to a known reference and shown to be 5 minutes slow. This is a calibration.

2) You may take a flight to a neighboring time zone and move your watch ahead one hour to compensate. This is an adjustment. No comparison to a known standard is involved, but this adjustment should get the watch closer to displaying the correct time.

3) You can compare the time on your watch to time.gov and then set your watch to match that time as closely as possible. This is both calibration and adjustment. Q-Lab’s Universal Calibrator (UC) system makes use of both calibrations and adjustments. When you use a UC20 smart sensor, for example, to ensure correct lamp irradiance output every 500 hours in your Q SUN, that represents a calibration - comparison of onboard irradiance sensors to the UC20 reference - followed by an adjustment of the Q-SUN’s irradiance measurement system. Earlier, in the process, Q-Lab compares output values from a new UC20 smart sensor to a NIST-traceable irradiance standard. This is a calibration of the UC20, followed by an adjustment of the UC20 so it can provide the best possible readings when the device is sent to you.

Szerző: Q Lab Blog
2022.08.12.

További híreink

What is a climatic chamber?

A climatic chamber, also known as an environmental chamber or climate chamber, is an enclosed space used to test the effects of specified environmental conditions on biological items, industrial
Read more >

UVC testing standard

ASTM International has recently announced the publication of ASTM G224, Standard Practice for Operating UVC Lamp Apparatus for Exposure of Materials. This is the first international standard to
Read more >

ISO 9227: How to conduct a salt spray test?

This methodology shall be used to perform the salt spray test standard according to ISO 9227-2017 in a corrosion chamber.  It should be used in conjunction with the ISO 9227-2017 standard document.
Read more >

Legfrissebb Hírek

What is a climatic chamber?

UVC testing standard

ISO 9227: How to conduct a salt spray test?

Outdoor exposure in Q-SUN and QUV


Értesüljön ön is elsőkézből!_eng

Amennyiben szeretne értesülni aktuális eseményeinkről, iratkozzon fel email címe és neve segítségével! Feliratkozásával hozzájárul adatai kezeléséhez. Adatait harmadik félnek nem adjuk ki.

SolverUnio Kft. All Rights reserved     Adatvédelem     Impressum